Residual magnetism demagnetization testing measures the remanent flux density of magnetized materials and quantifies how much magnetic flux is lost or removed through demagnetization processes. This testing is essential for verifying the long-term stability of permanent magnets and for confirming that ferromagnetic components meet strict residual field limits after demagnetization. Accredited third-party laboratories perform these evaluations in accordance with internationally recognized standards such as IEC 60404-5, ASTM A977, and GB/T 3217.
Technical Overview
Residual magnetism, also known as remanence, is the magnetization that remains in a ferromagnetic or permanent magnet material after an external magnetizing field is removed. It is quantified by the residual flux density Br, expressed in tesla (T) or millitesla (mT), or by the magnetic dipole moment for complete components. Demagnetization is the process of reducing or eliminating this residual magnetization, typically by applying a decaying alternating field, a reversed field pulse, or by heating the material above its Curie temperature.
Demagnetization amount testing has two distinct objectives. For permanent magnet materials, it quantifies irreversible flux loss caused by elevated temperature, opposing external fields, mechanical shock, or aging, which directly determines whether a magnet grade remains suitable for its operating environment. For machined steel parts, weldments, and workpieces that have undergone magnetic particle inspection, it verifies that the residual surface field has been reduced below specified limits so the part will not attract ferrous debris, disturb sensors, or interfere with subsequent machining, welding, or coating operations.
The testing framework is governed by international and national standards, including IEC 60404-5 and IEC 60404-8-1 for permanent magnet measurement and specification, IEC 60404-14 for dipole moment measurement by the withdrawal method, ASTM A977/A977M for high-coercivity permanent magnets using hysteresisgraphs, ASTM A341/A341M for permeameter methods, GB/T 3217 for permanent magnet magnetic property measurement, JIS C 2501 for permanent magnet test methods, and ASTM E1444/E1444M together with ISO 9934 for demagnetization acceptance after magnetic particle testing.
Testing Samples
Sintered and Bonded Permanent Magnet Materials
Test specimens cover the major commercial magnet families: sintered and bonded neodymium-iron-boron (NdFeB), samarium-cobalt (SmCo 1:5 and 2:17), hard ferrite (SrFe or BaFe based), and aluminum-nickel-cobalt (AlNiCo). Specimens are typically machined into standard geometries such as 10 mm × 10 mm × 10 mm blocks or Ø10 mm × 10 mm cylinders with ground, parallel end faces, as required for closed-circuit hysteresisgraph measurement. Bonded and injection-molded magnet grades can be tested in smaller or irregular geometries using open-circuit coil methods or vibration sample magnetometry.
Magnet Assemblies and Rotating Machine Components
Complete magnet assemblies, including motor rotor assemblies, Halbach arrays, sensor magnets, voice coil motors, magnetic couplings, magnetic chucks, and loudspeaker assemblies, are tested in their as-built condition. Because these samples cannot be placed in a closed magnetic circuit, their magnetic moment is measured with Helmholtz coils and fluxmeters, and the demagnetization amount is tracked as a percentage of the initial total flux.
Ferromagnetic Workpieces Requiring Demagnetization
Steel components that must carry low residual magnetism are tested after demagnetization processing. Typical samples include shafts, gears, bearings, crankshafts, camshafts, fasteners, springs, weldments, castings, and forgings, particularly those inspected by magnetic particle testing in accordance with ASTM E1444 or ISO 9934, which requires post-inspection demagnetization when residual fields could affect part function or subsequent processing.
Tooling, Fixtures, and Precision Instruments
Machining chucks, tool holders, clamping fixtures, precision gauges, and components used in semiconductor manufacturing, electron-beam systems, and analytical instruments are tested to confirm residual fields are below the low-millitesla or microtesla levels demanded by these sensitive environments.
Testing Items
The scope of residual magnetism and demagnetization amount testing typically includes the following parameters:
- Residual flux density Br (remanence) of magnet materials and assemblies
- Complete demagnetization curve, including second-quadrant B–H and intrinsic J–H curves
- Normal coercivity HcB and intrinsic coercivity HcJ
- Maximum energy product (BH)max and knee-point field Hk
- Magnetic dipole moment and total flux of assemblies measured by the withdrawal method
- Surface and air-gap residual field distribution mapped with Hall-effect gaussmeters
- Irreversible demagnetization loss percentage after thermal, field, or mechanical conditioning
- Demagnetization efficiency verification for AC, pulsed, and thermal demagnetization processes
- Temperature coefficients of Br and HcJ for operating point stability assessment
- Long-term aging and stability tests at specified operating temperatures
The demagnetization amount is calculated as the relative flux loss, η = (Φ0 − Φ1) / Φ0 × 100%, where Φ0 is the initial flux or magnetic moment and Φ1 is the value measured after the demagnetizing condition has been applied and the sample has stabilized at room temperature. For demagnetized workpieces, the result is reported as a residual surface field strength in mT or gauss, compared directly against the applicable acceptance limit.
Testing Methods
Closed-Circuit Hysteresisgraph Testing (IEC 60404-5, ASTM A977, GB/T 3217)
The specimen is clamped between the pole shoes of a high-field electromagnet to form a closed magnetic circuit, with Hall probes positioned at the pole faces. The sample is first pulse-magnetized to full saturation, then the demagnetization curve is traced through the second quadrant of the hysteresis loop. This method directly yields Br, HcB, HcJ, (BH)max, and the full B–H/J–H curves, and provides the reference baseline flux used to evaluate subsequent demagnetization loss. Specimens must be dimensionally accurate, with flat and parallel pole contact faces, and are conditioned at 23 °C ± 5 °C before measurement.
Open-Circuit Flux Measurement with Helmholtz Coil and Fluxmeter (IEC 60404-14)
For magnets that cannot be tested in a closed circuit, the sample is centered in a calibrated Helmholtz coil pair and withdrawn or rotated while an integrating fluxmeter records the linked flux, from which the magnetic dipole moment is derived. This technique is ideal for assembled rotors, ring magnets, and irregular components. Combined with a pulsed magnetizer or demagnetizer, it enables direct measurement of the demagnetization amount by comparing the magnetic moment before and after each conditioning step.
Vibration Sample Magnetometry (VSM)
Small specimens, millimeter-scale coupons, and research samples are characterized on a vibration sample magnetometer, in which the sample oscillates within detection coils under a swept applied field of up to approximately 2–3 T, or higher with superconducting magnets. VSM is particularly suitable for high-coercivity NdFeB grades, thin films, bonded magnet powders, and for determining temperature-dependent demagnetization behavior with furnace or cryostat options.
Surface and Ambient Residual Field Measurement
Residual magnetism on demagnetized parts is measured with a calibrated Hall-effect gaussmeter, scanning the probe across defined grid points or performing a peak-field search over the part surface. Fluxgate and magnetoresistive magnetometers are applied when verification down to the 0.01 mT level is required. For magnetic particle testing acceptance, ASTM E1444 commonly uses a residual field criterion of no more than 3 gauss (0.3 mT) at the part surface unless the product specification dictates otherwise.
Demagnetization Amount Test Procedure
A representative test sequence is as follows:
- Saturate the magnet and record the initial flux Φ0 or magnetic moment at reference temperature
- Expose the sample to the specified demagnetizing condition, such as thermal aging at 80 °C to 200 °C per the temperature class, reverse-field pulses at defined amplitudes, a decaying 50/60 Hz AC field, or mechanical shock and vibration profiles
- Allow the sample to stabilize at room temperature for a defined holding period
- Re-measure the flux Φ1 and compute the irreversible loss percentage
- Where applicable, repeat the measurement after re-exposure to the conditioning temperature to separate the reversible loss component from the total loss
Typical reference values illustrate the scale of the parameters involved: a sintered NdFeB grade N42 magnet exhibits Br of approximately 1.28–1.32 T with HcJ of at least 955 kA/m, sintered SmCo 2:17 grades exhibit Br of about 1.05–1.12 T, and hard ferrites show Br of roughly 0.39–0.42 T. NdFeB temperature coefficients of approximately −0.11 to −0.13 %/°C for Br and −0.5 to −0.65 %/°C for HcJ explain why irreversible loss testing is critical for elevated-temperature applications.
Applications
Electric Motors, EV Traction Drives, and Generators
Demagnetization testing validates that rotor magnets withstand armature reaction fields, overload transients, and operating temperatures without exceeding allowable irreversible flux loss, which is decisive for torque density and efficiency retention in traction motors, servos, and wind turbine generators.
Aerospace and Defense Systems
Actuators, fuel system components, and magnetically sensitive avionics require quantified magnet stability across thermal cycling, shock, and vibration profiles, and often demand verified demagnetization of structural parts to prevent magnetic signature or debris attraction.
Sensors and Precision Instruments
Encoders, flow meters, Hall sensor modules, and precision balances depend on stable reference fields; residual magnetism testing confirms both magnet stability and the absence of stray fields from housings and fixtures that could bias readings.
Manufacturing and Post-Inspection Quality Control
Components inspected by magnetic particle testing, and parts processed through magnetic chucks, lifting magnets, or magnetic separation, are routinely verified for residual field compliance before assembly, plating, welding, or shipment, protecting bearings, gears, and hydraulic components from ferrous debris contamination.
Electronics, Semiconductor, and Medical Equipment
Low residual magnetism levels are specified for electron-beam lithography fixtures, vacuum systems, analytical instruments, and implantable or MRI-adjacent devices, where even weak stray fields can distort beam trajectories or imaging performance.
FAQ
What is the difference between remanence Br and residual magnetism on a component?
Remanence Br is a material property measured under closed-circuit conditions on a standard specimen and represents the flux density remaining after saturation. Residual magnetism on a component refers to the stray field present on a real part in its working geometry, which depends on the material, shape, permeance coefficient, and magnetization history. The two are related but are measured with different instruments and reported in different terms.
What demagnetization level is acceptable after magnetic particle inspection?
Under ASTM E1444 and common industrial practice, a residual field of 3 gauss (0.3 mT) or less at the part surface is generally considered acceptable when the specification does not define a stricter limit. Certain applications, such as bearings or components for sensitive instruments, impose limits down to 0.5 gauss or lower.
How are reversible and irreversible demagnetization losses distinguished?
Irreversible loss is the flux reduction that persists after the magnet returns to room temperature and is calculated from the before-and-after flux measurements. Reversible loss is the flux change present only while the magnet is at elevated temperature and is recovered on cooling; it is assessed by re-heating the sample and remeasuring, or derived from the temperature coefficient of Br.
What magnetizing field is required to saturate an NdFeB sample before testing?
The applied field should reach at least three times the intrinsic coercivity HcJ, which for high-coercivity NdFeB grades typically means magnetizing fields of 2,400 kA/m (approximately 3 T) or more, delivered by pulsed capacitor-discharge magnetizers or high-field electromagnets.
Can assembled magnets or complete rotors be tested without disassembly?
Yes. Open-circuit Helmholtz coil measurements per IEC 60404-14 determine the magnetic moment of the complete assembly, and the demagnetization amount is evaluated by comparing repeated measurements after each thermal or field conditioning cycle. Surface field mapping with a gaussmeter provides supplementary spatial verification.
How are results reported and what measurement uncertainty applies?
Reports issued by ISO/IEC 17025 accredited third-party laboratories include the measured Br, HcB, HcJ, (BH)max, curve data, demagnetization loss percentages at each test condition, residual field maps where applicable, instrument calibration traceability, and expanded uncertainty, which is typically on the order of 1–3% for hysteresisgraph flux density measurements.